 Shack-Hartmann wavefront sensor test of single lens 
Setup:
In the above configuration, diverging light from Optino falls on the lens L under test. It is made parallel by it, and falls on a flat mirror M (which should be of a high quality, say λ/4, λ/10 or λ/20, depending on the accuracy required for testing the lens L). The light from the mirror M returns to Optino, and passed to the SH system.

The aberrations of the SH system itself are removed taking a calibration frame of a small high quality (like for the flat mirror above) spherical mirror instead of the lens L.

The mirror should have a minimum diameter equal to that of the lens L, and the focal ratio of the lens C2 should match that of the lens L.
download test configuration of single lens with flat mirror